350 Cable Survey System
DPN 402197 © TSS (International) Ltd Page 10 of 30
❐ The field +1234 represents a value of 1.23 × 10
4
microvolts (or 12.3 mV). The
SDC would display this on the Run Display screen as +1.23e4 in the lower
left-hand data panel.
❐ The field +2416 represents a value 2.41 × 10
6
microvolts (or 2.41 volts). The
SDC would display this on the Run Display screen as +2.41e6 in the lower
left-hand data panel.
10. The QC check code provides additional status information that explains any
occurrence of the QC flag being set. The check code consists of a two-digit
number in the range 01 to 07 and 99 with the meanings defined in Table 7–2.
7.3.1.2 Forward Search mode
The string is 48 characters long with individual field definitions as follows. The SDC
logs all distances in units of centimetres and signal voltages in units of microvolts
using the scientific notation. The values in the packet are rounded and it is possible
that they will not precisely match those on the Forward Search screen.
Table 7–3: External logging format – Forward search mode
Table 7–2: QC check code meaning – Survey mode
QC Check
Code
Meaning
00 Target in range. SL and PL ≥50µV; LAT ≤±2m. Quality flag is RESET.
01 Target in range. SL or PL <50µV; LAT ≤±2m. Quality flag is SET.
02 Target in range. SL or PL ≥50µV; LAT >±2m. Quality flag is SET.
03 Target in range. SL or PL <50µV; LAT >±2m. Quality flag is SET.
04 Starboard tracking data only; LAT = ????, VRT = ????, SKEW = ???. Quality flag is SET.
05 Port tracking data only; LAT = –????, VRT = ????, SKEW = ???. Quality flag is SET.
06 Skew angle not available. Skew angle = ???. Quality flag is SET.
07 Saturation in one or more coils. Quality flag is SET. The SDC displays a warning banner on the Run Dis-
play screen.
99 Target out of range. VRT and LAT = ????. SKEW = ???. Quality flag is SET.