9 – Maintenance
DPN 402197 © TSS (International) Ltd Page 17 of 26
9.3.1 Fault on a Single Channel
Figure 9–7: Single channel failure
Power-off the System
Swap the coil cable
on the faulty side.
Swapthecoilonthe
faulty channel
Is the
channel
working?
Yes
No
No
Yes
Is the
channel
working?
Enter new calibration
details into SDC.
See sub-section 6.2.2.
Is the
wiring
good?
Is the
channel
working?
Contact TSS for advice
No
Yes
No
Yes
Single channel fault
Check connections and
wiring to the ADC Board
Disassemble the SEP.
See sub-section
9.2.2
Resume the survey
Renew the faulty
coil cable
Renew the faulty coil.
See sub-section 3.3.2.
Repair or renew the
connector wiring.
Renew the ADC Board